Optris Xi 400 LT
Optris

Xi 400 LT compact fixed thermal imaging camera with USB

Pyrometers & thermal cameras

Manufacturer
Optris
Model
Xi 400 LT
Instrument type
Pyrometers & thermal cameras
Detector
Uncooled bolometer, 382 × 288 pixels, 17 µm pitch
Spectral range
8–14 µm
Frame rate
80 Hz (switchable to 27 Hz)
Lenses (FOV)
18° × 14°, 29° × 22°, 53° × 38°, 80° × 54°
Distance-to-spot ratio
390:1, 239:1, 128:1, 78:1 (by lens)
Temperature ranges
−20–100 °C, −20–900 °C, 200–1500 °C (−20–1500 °C on some lenses)
Accuracy
±2 °C or ±2 %, whichever is greater
Thermal sensitivity (NETD)
50 mK
Focus
Motorised
Interface
USB 2.0; optional USB GigE (PoE)
Direct I/O
1 × analogue in 0–10 V, 1 × digital in, 1 × analogue out 0–10 V
Process interface (option)
2 AI, 1 DI, 3 × 0/4–20 mA outputs or relays, fail-safe relay
Housing
Stainless steel, Ø 36 × 100 mm, M30 × 1 thread
Protection
IP67, NEMA 4
Operating temperature / power
0–50 °C; via USB, max. 2.5 W
Request a quote
Review enquiry

The Xi 400 LT is a compact industrial fixed thermal imaging camera for non-contact infrared temperature measurement. It is connected and powered via USB and delivers radiometric thermal images that provide detailed temperature profiles for industrial monitoring, quality control and safety tasks.

The camera works in the long-wave infrared range of 8 to 14 µm with an uncooled two-dimensional bolometer of 382 × 288 pixels and 17 µm pixel pitch. Images are captured at 80 Hz, switchable to 27 Hz. Thermal sensitivity (NETD) is 50 mK. A motorised focus and distance-to-spot ratios of up to 390:1 allow small objects to be measured precisely from a distance. Emissivity, transmissivity and reflectivity are adjustable from 0.100 to 1.100.

Four lenses are available: 18° × 14° (focal length 20 mm, D:S 390:1), 29° × 22° (12.7 mm, 239:1), 53° × 38° (7.7 mm, 128:1) and 80° × 54° (5.4 mm, 78:1). The smallest detectable spot (one pixel) is 300 µm to 800 µm and the smallest measurable spot (3 × 3 pixels) 0.9 mm to 2.4 mm, depending on lens; minimum target distance is 200 to 350 mm. Each lens can be combined with the temperature ranges −20 to 100 °C, −20 to 900 °C and 200 to 1500 °C (−20 to 1500 °C for some lenses). Accuracy is ±2 °C or ±2 %, whichever is greater; warm-up time is 10 minutes.

The camera has a USB 2.0 interface with cables of 1 m as standard and 3, 5, 10 or 20 m as options; an optional USB GigE (PoE) interface is offered. Directly on the camera there is one analogue input (0–10 V), one digital input (max. 24 V) and one analogue output (0–10 V). An optional industrial process interface adds two analogue inputs, one digital input, three 0/4–20 mA outputs or relays (0–30 V / 400 mA) and a fail-safe LED and relay. Power consumption is at most 2.5 W, supplied via USB.

The stainless steel housing measures Ø 36 × 100 mm with an M30 × 1 thread and 1/4-20 UNC tripod mount and weighs 216–220 g depending on lens. It is rated IP67 / NEMA 4, operates from 0 to 50 °C and is tested for shock to IEC 60068-2-27 and vibration to IEC 60068-2-6 and -64; it carries CE and UKCA marking and is RoHS compliant. A mounting bracket is included, and a water-cooled housing and air purges are available. Operating software is required for use.

Applications named include early fire detection, electrical substation monitoring, fever screening, injection moulding, PCB development, food temperature measurement and cold-chain monitoring.

Please state the required lens, temperature range, interface and cable length in your enquiry.

Part numbers · 36

OPTXI40LTF20T010, OPTXI40LTF20T090, OPTXI40LTF20T150, OPTXI40LTF13T010, OPTXI40LTF13T090, OPTXI40LTF13T150, OPTXI40LTF08T010, OPTXI40LTF08T090, OPTXI40LTF08T150, OPTXI40LTF06T010, OPTXI40LTF06T090, OPTXI40LTF06T150, DDPXI40LTF20T090, DDPXI40LTF13T090, DDPXI40LTF08T090, DDPXI40LTF06T090, DDPXI40LTF20T010, DDPXI40LTF13T010, DDPXI40LTF08T010, DDPXI40LTF06T010, DDPXI40LTF20T150, DDPXI40LTF13T150, DDPXI40LTF08T150, DDPXI40LTF06T150, MEPXI40LTF20T090, MEPXI40LTF13T090, MEPXI40LTF08T090, MEPXI40LTF06T090, MEPXI40LTF20T010, MEPXI40LTF13T010, MEPXI40LTF08T010, MEPXI40LTF06T010, MEPXI40LTF20T150, MEPXI40LTF13T150, MEPXI40LTF08T150, MEPXI40LTF06T150

Industry applications

Selection notes

Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.