Fluke 1587 FC
Fluke

1587 FC 2-in-1 insulation tester and true-RMS multimeter

Multimeters & clamp meters

Manufacturer
Fluke
Model
1587 FC
Instrument type
Multimeters & clamp meters
Insulation resistance
0.01 MΩ to 2 GΩ
Insulation test voltages
50, 100, 250, 500, 1000 V
AC voltage
30.0 mV to 1000 V, true-RMS
DC voltage
1 mV to 1000 V
AC/DC current
0.01 mA to 400 mA
Resistance
0.1 Ω to 50.0 MΩ
Capacitance
1 nF to 9999 µF
Temperature
−40 to +537 °C (K-type)
Display
6000 counts, 3¾ digits
Special functions
VFD low-pass filter, PI/DAR, live circuit detection > 30 V, auto-discharge
Wireless
2.4 GHz radio, mobile app
Safety rating
CAT IV 600 V / CAT III 1000 V (IEC 61010-2-033)
Batteries
4 × AA; approx. 1000 h meter use or ≥ 1000 insulation tests
Operating temperature
−20 to 55 °C
Size / weight
5.0 × 10.0 × 20.3 cm; 550 g
Request a quote
Review enquiry

The 1587 FC is a handheld 2-in-1 instrument that combines an insulation tester with a true-RMS digital multimeter. It is designed for electrical troubleshooting and preventive maintenance and is used by solar professionals, marine electricians including on navy ships, motor shops and technicians working on HVAC compressor motors.

For insulation testing the instrument sources a test voltage of 50, 100, 250, 500 or 1000 V and measures insulation resistance from 0.01 MΩ to 2 GΩ. Live circuit detection prevents a test if a voltage above 30 V is present, and capacitive voltage is discharged automatically after the test. Timed polarisation index (PI) and dielectric absorption ratio (DAR) tests are performed with trend graphs on a smartphone, and temperature compensation of insulation readings is supported.

The multimeter section measures AC voltage from 30 mV to 1000 V and DC voltage from 1 mV to 1000 V, AC/DC current from 0.01 mA to 400 mA, resistance from 0.1 Ω to 50 MΩ, capacitance from 1 nF to 9999 µF and temperature from −40 to +537 °C, plus diode and continuity tests. AC measurements are true-RMS. A VFD low-pass filter allows accurate measurements on motor drives. The display has 6000 counts (3¾ digits).

The meter communicates over a 2.4 GHz radio link with a mobile app to share measurements, store data and monitor remotely. It is powered by four AA batteries, giving about 1000 hours of meter use or at least 1000 insulation tests of 1000 V into 1 MΩ. The mA input is fuse-protected (0.44 A, 1000 V, 10 kA).

The safety rating is CAT IV 600 V / CAT III 1000 V according to IEC 61010-2-033, with CSA and TÜV certification. Operating temperature is −20 to 55 °C, operating altitude up to 2000 m and enclosure protection IP40. EMC conforms to IEC 61326-1 and -2-2 and CISPR 11, and the radio holds FCC and IC certification. The instrument measures 5.0 × 10.0 × 20.3 cm and weighs 550 g.

The FLK-1587-FC set includes a remote probe, test leads, alligator clips, a K-type thermocouple, soft case and documentation. Accessories include magnetic hangers, an i400 AC current clamp and a C25 soft case. Related models are the 1587 without wireless connectivity, the 1587T with 50 V and 100 V test voltages, and the 1577 with 500 V and 1000 V tests from 0.1 MΩ to 600 MΩ.

Please state the required model, accessories and quantity in your enquiry.

Part numbers · 3

FLK-1587-FC, 1587 FC, 1577

Industry applications

Selection notes

Match the instrument to the circuit and measurement category. Include the wiring arrangement and ratios of any external current or voltage transformers.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.