Type 8203
Liquid analysis

pH & ORP electrodes
The AH-151 is an industrial pH meter with automatic calibration and self-cleaning functions, designed for continuous monitoring of process water. By carrying out electrode cleaning and calibration automatically, it reduces the maintenance workload.
The pH value is measured by the glass electrode method, together with the temperature of the water. At set intervals the system automatically cleans the electrode chemically and calibrates it with standard solutions. The solutions are atomised, which reduces their consumption to one fifth or less. The cleaning interval can be set from 0.1 to 168 hours and the calibration interval from 1 to 999 hours. Self-diagnostic functions monitor the condition of the electrode, and the electrode characteristics are displayed graphically.
The measuring range is pH 0 to 14 and 0 to 100 °C, with resolutions of 0.01 pH and 0.1 °C. Repeatability is within ±0.02 pH and ±0.3 °C, and linearity within ±0.03 pH and ±0.3 °C.
Operation and display are via a 5.7-inch colour TFT touch panel. Measured values are transmitted as 4–20 mA DC signals; there are nine contact outputs plus one abnormality alarm output, three contact inputs and an RS-485 two-wire communication interface.
The instrument is built for outdoor installation with a splash-proof structure to JIS C0920 (IP54 equivalent). The control unit case is SPCC steel coated with melamine-based epoxy resin; the metal fittings and stand are SUS304. Apart from the mounting bracket and the lifting unit, all parts are integrated into a single unit, which simplifies installation. The operation unit with pole stand weighs about 60 kg, the metal fittings about 3 kg and the lifting unit about 1.5 kg. Power is 100–240 V AC ±10 %, 50/60 Hz, with about 20 W consumption in steady state and 35 W during automatic calibration. The operating temperature range is 0 to 50 °C. The device conforms to RoHS, REACH and EMC requirements.
The AH-151-CH is the dip-type version; flow-through (AH-151-CF) and water sampling (AH-151-FK) versions are also listed.
Applications named include process water, drinking water utilities, water reuse, water testing and waste water treatment.
Please state the required installation type (dip, flow-through or sampling), the medium and the quantity when you enquire.
AH-151-CH
Record the sample temperature, expected concentration and how the probe is mounted. Identify the mating transmitter when replacing a sensor.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.