CTlaser MT infrared pyrometer for measuring through flames

Pyrometers & thermal cameras

Manufacturer
Optris
Model
CTlaser MT
Instrument type
Pyrometers & thermal cameras
Temperature range
200–1450 °C (MT); 400–1650 °C (MTH)
Spectral range
3.9 µm (MT); 4.24 µm (MTH)
Optical resolution
45:1
Response time
300 µs (MT); 10 ms (MTH)
Accuracy
±1 % (MT); ±1.5 °C or ±1 % (MTH)
Repeatability
±0.5 % of reading + 0.5 °C (MT); ±0.5 °C or ±0.5 % (MTH)
Optics
standard focus SF; close focus CF1–CF4 (1.6 mm at 70 mm to 10 mm at 450 mm)
Sighting
double laser, 635 nm, max. 1 mW
Emissivity
adjustable 0.100–1.100
Interfaces
USB, RS-232, RS-485, Profibus DP, Ethernet, relay; analogue outputs
Protection
IP65 (NEMA 4)
Request a quote
Review enquiry

CTlaser MT is an industrial infrared pyrometer designed for a task where ordinary pyrometers fail: measuring the temperature of a surface behind a flame. Hot combustion gases emit strongly in most of the infrared band, so a broadband sensor reads the flame instead of the product. The MT works at 3.9 µm and the MTH at 4.24 µm, narrow spectral windows chosen to suppress flame radiation, so the reading reflects the steel, glass or other material being heated.

Two versions cover different temperatures. The MT measures from 200 to 1450 °C with a very fast response time of 300 µs, suited to flame hardening and moving products. The MTH measures from 400 to 1650 °C with a response time of 10 ms. Accuracy is ±1 % for the MT and ±1.5 °C or ±1 % for the MTH; thermal sensitivity is 180 mK and 100 mK respectively.

The pyrometer is a two-piece design: a sensing head with optics and a separate electronics box, which allows the head to be installed close to the process while the electronics stay in a cooler place. Optical resolution is 45:1. Standard-focus optics give a 26.7 mm spot at 1200 mm, and four close-focus options produce spots from 1.6 mm at 70 mm to 10 mm at 450 mm. A double laser marks the measured spot exactly, which simplifies alignment on small targets.

Signal processing includes adjustable emissivity and transmissivity, averaging, peak and valley hold. Outputs and interfaces can be selected from analogue signals, USB, RS-232, RS-485, Profibus DP, Ethernet or relay modules, so the instrument fits existing control systems. The head is IP65 rated; for hot surroundings cooling jackets, air purge collars, furnace wall mounts and high-temperature cables up to 180 °C are offered.

Typical applications are flame hardening, furnace and combustion monitoring in metal processing, float glass and glass tube production, thermoforming and plastics, brake disc and tyre testing, semiconductor and solar processes, battery cell sealing and early fire detection. Configuration software and a mobile app support setup and diagnostics. Tell us the target temperature, distance and spot size, and we will propose the matching version and optics.

Part numbers · 4

OPTCTLMT, OPTCTLMTH, MT, MTH

Industry applications

Selection notes

Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU