HORIBA ELCHL / CHLSET
HORIBA

ELCHL amperometric residual chlorine sensor, 0–20 mg/l

Dissolved oxygen, chlorine & ozone sensors

Manufacturer
HORIBA
Model
ELCHL / CHLSET
Instrument type
Dissolved oxygen, chlorine & ozone sensors
Measured parameter
residual chlorine (free or total)
Measuring principle
membrane-covered, three-electrode amperometric
Measuring range
0–20 mg/l (ppm)
Accuracy
±0.1 mg/l
Repeatability / reproducibility
±0.05 mg/l (25 °C) / ±4 %
Resolution
0.01
Response time
T90 90 s at 25 °C
Sample pH
4 to 12
Operating temperature
0 to 50 °C, automatic temperature compensation
Operating pressure
max. 15 psi
Flow requirement
30 to 60 l/h (0.6 to 1.3 cm/s)
Materials
body PVC, ABS; membrane hydrophilic PTFE; electrodes silver–silver halide, gold, 316 SS
Articles
ELCHL sensor; ELCHL-M membrane cap; ELCHL-R electrolyte; ELPH pH electrode; CHLSET complete set
Request a quote
Review enquiry

This chlorine sensor measures residual chlorine, free or total, in water. It is intended for industrial and municipal waste water treatment plants, drinking water treatment plants, surface water applications and other process water applications. The sensor is supplied as the ELCHL electrode alone or as part of the CHLSET complete measuring set.

The sensor works on membrane-covered, three-electrode amperometric technology. Chlorine diffuses through a hydrophilic PTFE membrane to the electrodes, where it produces a current proportional to its concentration. The electrodes are silver–silver halide, gold and 316 stainless steel. The membrane cap is replaceable and the gel electrolyte is refillable. Temperature compensation is automatic, and the sensor is factory calibrated, so no zero calibration is required.

The measuring range is 0–20 mg/l (ppm) with a resolution of 0.01. Accuracy is ±0.1 mg/l, repeatability ±0.05 mg/l at 25 °C and reproducibility ±4 %. The response time T90 is 90 seconds at 25 °C. The sensor works at sample pH from 4 to 12, at operating temperatures from 0 to 50 °C and at operating pressures up to 15 psi.

Amperometric sensors depend on a steady flow of sample past the membrane. For this sensor the flow should be between 30 l/h (0.6 cm/s) and 60 l/h (1.3 cm/s).

The sensor body is made of PVC and ABS. The product complies with RoHS, REACH and EMC requirements.

Several articles are available. ELCHL is the amperometric chlorine electrode only. ELCHL-M is the replacement membrane cap and ELCHL-R the refillable electrolyte for the ELCHL. ELPH is a general-purpose online pH electrode. CHLSET is the complete amperometric chlorine set, which includes the chlorine electrode, a pH electrode, a holder, a flow meter, a mounting plate and a 3-foot cable. Because the chlorine reading can be evaluated together with pH, the set provides both measurements at one installation.

To request a quotation, please state whether you need the complete CHLSET or individual items, the expected chlorine range and the number of measuring points.

Part numbers · 5

CHLSET, ELCHL, ELCHL–M, ELCHL–R, ELPH

Industry applications

Selection notes

Record the sample temperature, expected concentration and how the probe is mounted. Identify the mating transmitter when replacing a sensor.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU