Type 8203
Liquid analysis

Water quality analysers
TOCADERO ONE is an automatic online analyser for total organic carbon (TOC). It is designed for precise and reliable water quality monitoring in industrial and environmental applications, in particular wastewater, where samples can contain high concentrations of organic carbon and dissolved salts.
The analyser oxidises the sample by combustion at 1200 °C without a catalyst. This allows high-concentration samples, up to 20,000 mg/l C, to be handled. The resulting carbon dioxide is carried by CO2-free instrument air to the detection system. The analyser determines TC and NPOC; a TC/TOC result is available in 1 minute and a TOC/NPOC result in less than 3 minutes. Condition-based maintenance uses sensor monitoring to plan service, and the system cleans itself automatically with pure water.
Measuring ranges are 0–1, 0–100, 0–1,000 and 0–10,000 ppm, with an optional range of 0–100,000 ppm. The limit of detection is 0.1 ppm and repeatability is ±2 % of end of range. The analyser accepts total dissolved solids up to 200 g/l (20 % NaCl) and particles below 2000 µm; a homogeniser and sample preparation are available as options.
Operation is via a 7-inch touch panel. Four programmable relays to NAMUR standard are provided. Communication supports Profinet, Modbus and OPC UA over Ethernet, and wireless options include Bluetooth, WLAN, GSM/LTE and 5G. Power supply is 110–230 V AC ±10 V, 50/60 Hz, approximately 600 VA.
The housing is rated IP54 or IP65, with a NEMA 4X option. For hazardous areas an Ex p enclosure is available for Zone 1 and Zone 2, temperature classes T3 and T4, with ATEX and IECEx approval. The analyser measures 800 × 600 × 320 mm (H × W × D) and weighs less than 55 kg. It is installed indoors at 2 to 45 °C and below 85 % relative humidity without condensation, and complies with RoHS, REACH and EMC. On-site support is available.
Applications include wastewater management, sewage treatment plants, factory wastewater analysis, detection of organic contamination in condensate and monitoring of TOC levels in airport wastewater.
Please describe the sample type, expected TOC range and salt content, and whether an Ex p version is needed.
Record the sample temperature, expected concentration and how the probe is mounted. Identify the mating transmitter when replacing a sensor.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.